title = {Imaging Radiation Dose Reduction with Adaptive Statistical Iterative
Reconstruction in Brain CT},
journal = {Journal of Engineering and Applied Sciences},
volume = {12},
number = {24},
pages = {7431-7434},
year = {2017},
issn = {1816-949x},
doi = {jeasci.2017.7431.7434},
url = {https://makhillpublications.co/view-article.php?issn=1816-949x&doi=jeasci.2017.7431.7434},
author = {Cheong-Hwan,Yoon-Jeong,Dan-Bi,Ino,Su-Chul and},
keywords = {dose reduction,ASIR (Adaptive Statistical Iterative Reconstruction Algorithm),FBP (Filtered Back Projection),patient,quantitative,equivalence},
abstract = {We compared and analyzed patient doses depending on the application of FBP (Filtered Back
Projection) and ASIR (Adaptive Statistical Iterative Reconstruction Algorithm). A64-MDCT Optima ™ CT 660
system was used for imaging. For quantitative analysis, images containing both sides of orbit and basal ganglia
were obtained. After setting a 40~45 mm2 area as Region of Interest (ROI), the average CT number was
measured. A Statistical Program, SPSS 22.0 was used for the calculations and ANOVA, Kruskal-Wallis and
paired t-tests were performed after the normality and equivalence tests. For quantitative analysis, a comparison
of SNR of each scan type was carried out. From the Kruskal-Wallis test, the SNR of axial scan was 0.13±0.67
and the SNR of Helical scan was 0.12±0.64. There was a significant difference for the axial scan values
(χ2= 12.54, p<0.05) but there was no corresponding significant difference for those from the Helical scan
(χ2 = 0.49, p>0.05). When ASIR was applied in the axial scan, the value of CTDIvol
decreased by 21.46 mGy
(47.8%) at maximum. The value of DLP decreased by 343.42 mGy•cm (47.8%) at maximum. When ASIR was
applied in the Helical scan, the value of CTDIvol
decreased by 18.52 mGy (49.9%) at maximum and the value of vol
DLP decreased by 328.92 mGy•cm (49.9%) at maximum. This study could confirm a dose reduction through
applying FBP and ASIR algorithm for each scan type in brain CT imaging.}
}