TY - JOUR
T1 - CO2 Laser Photoacoustic for Measuring NH3 Gas Concentration on Exhaled Breath of Scavengers on Landfill
AU - Oktafiani, Fitri AU - Stiyabudi, Rizky AU - Hendra Gunawan, Wahyu AU - , Mitrayana
JO - Journal of Engineering and Applied Sciences
VL - 12
IS - 14
SP - 3687
EP - 3691
PY - 2017
DA - 2001/08/19
SN - 1816-949x
DO - jeasci.2017.3687.3691
UR - https://makhillpublications.co/view-article.php?doi=jeasci.2017.3687.3691
KW - Spectro meter
KW -scavengers
KW -photoacoustic
KW -occured
KW -concentration
AB - A CO2 laser photoacoustic spectrometer intracavity configuration has been built and tested. The
application was for determining the NH3 gas concentration on exhaled breath of scavengers on the Landfill. The
off-line experiment was conducted through non-invasive sample method using exhaled breath. Exhaled breath
sample was taken by three daily variations of the morning, mid-day and post-day. The data was analized by
multicompound gas method using the software of excel, origin and scilab. The result from the photoacoustic
spectrometer intracavity configuration with the strongest frequency resonance(1650±5) Hz at biomarker gas
occured on absorption lines 10R14, 10P14 and 10P16. The lowest detection boundary from ammonia is
(23.3±0.5) ppbV. The highest ammonia concentration from scavengers was (2904±1) ppbV occured in the early
morning and the lowest concentration was (1089±1) ppbV in the post-day. Meanwhile, on healthy volunteers,
the highest ammonia concentration was (1676±1) ppbV in the morning and the lowest ammonia concentration
was 377±1) ppbV in the mid-day. It shows the contamination of air ambien caused the increasing concentration
of biomarker gas in the human body.
ER -