files/journal/2022-09-02_12-54-44-000000_354.png

Journal of Engineering and Applied Sciences

ISSN: Online 1818-7803
ISSN: Print 1816-949x
104
Views
0
Downloads

Development Research Methodology Elastic Deformation Total Station

D.A. Gura, G.G. Shevchenko and A.Yu. Gura
Page: 2885-2888 | Received 21 Sep 2022, Published online: 21 Sep 2022

Full Text Reference XML File PDF File

Abstract

The results of experimental measurements with use of a electronic tacheometer (total station) proving the presence of elastic deformations of the device support caused by imbalance of the moving parts (the alidade). Deformations lead to errors in a measurement of horizontal angles. The study presents the research methods of the fixed part of a electronic tacheometer as well as ways to increase in several times the accuracy of determining the azimuth and vertical deformation of the support. For measurements use a special device with two optical mirrors located vertically at an angle of 90° to each other and fixed to the stationary part of the electronic tacheometer below the alidade but above the support. If there are no elastic deformations, then upon rotating the alidade of the horizontal circle of the tested electronic tacheometer, the device must be static. If elastic deformations exist, the device with mirrors may turn around the vertical axis (azimuthal rotations) that can be registered by other electronic tacheometers (measuring tacheometers). These turns lead to errors in measuring horizontal angles.


How to cite this article:

D.A. Gura, G.G. Shevchenko and A.Yu. Gura. Development Research Methodology Elastic Deformation Total Station.
DOI: https://doi.org/10.36478/jeasci.2016.2885.2888
URL: https://www.makhillpublications.co/view-article/1816-949x/jeasci.2016.2885.2888