files/journal/2022-09-02_12-54-44-000000_354.png

Journal of Engineering and Applied Sciences

ISSN: Online 1818-7803
ISSN: Print 1816-949x
105
Views
1
Downloads

Designing of Arrythmia Telemonitor Device Using Pulse Sensor Based on Web

R. Wahyu Kusuma, I. Ridha, E.P. Swelandiah and R. Yasman
Page: 1226-1232 | Received 21 Sep 2022, Published online: 21 Sep 2022

Full Text Reference XML File PDF File

Abstract

This research aims to design a prototype medical device that is expected to be applied for independent monitoring of the remote by the patient as well as to record medical data of cardiac arrhythmias. This research was conducted by using experimental model research with laboratory test. The arrhythmia telemonitor device that has been constructed is composed of pulse sensor, keypad and press switch as input block. Process block consisting of Arduino Mega Kit module, Ethernet Shield, router and USB Modem. The output block is composed of a 16×2 alpha numeric LCD viewer module and a web display accessed using a computer or smartphone device. Based on test results on five respondents obtained level of accuracy is based on the calculation of procentstase error average of 0.25%. While the results of precision testing based on the calculation Standard deviation (Stdev) averaged 13.27.


How to cite this article:

R. Wahyu Kusuma, I. Ridha, E.P. Swelandiah and R. Yasman. Designing of Arrythmia Telemonitor Device Using Pulse Sensor Based on Web.
DOI: https://doi.org/10.36478/jeasci.2019.1226.1232
URL: https://www.makhillpublications.co/view-article/1816-949x/jeasci.2019.1226.1232